The Notified Body review process represents the critical gateway to CE marking under the Medical Device Regulation. Having worked with numerous manufacturers through this process, we have observed that the quality and completeness of the initial submission directly correlates with certification timeline success.
This guide provides a systematic checklist approach to technical file preparation, organized according to MDR Annex II requirements. Use it as a final verification tool before submission to ensure nothing is overlooked.
Pre-Submission Fundamentals
Before diving into specific documentation requirements, ensure these foundational elements are in place:
Document Management System
- Version control: Every document should have a clear version number, date, and revision history.
- Document index: A master list of all technical file contents with document numbers, titles, versions, and locations.
- Approval status: All documents should show evidence of review and approval by qualified personnel.
- Cross-reference consistency: When Document A references Document B, verify the reference is accurate and the versions align.
Submission Format
- Notified Body requirements: Confirm your NB's specific format requirements (electronic submission portal, file naming conventions, etc.).
- Searchable PDFs: All documents should be text-searchable, not scanned images.
- Logical folder structure: Mirror the Annex II structure in your submission organization.
- File size limits: Verify maximum file sizes and split large documents appropriately.
Section 1: Device Description and Specification
Device Identification Checklist
- Product name and variants: Complete list of all product names, models, and variants covered by the submission.
- Basic UDI-DI: Obtained from an accredited issuing entity.
- Device classification: Classification rationale with reference to specific MDR Annex VIII rules applied.
- Intended purpose statement: Clear, specific statement aligned with all labeling and promotional materials.
- Intended users: Specification of who will use the device (healthcare professionals, lay users, etc.).
- Patient population: Target patient groups including any contraindications.
Technical Specifications Checklist
- Functional description: How the device works, including diagrams and illustrations.
- Key performance characteristics: Measurable parameters that define device function.
- Materials in contact with patient: Complete list with biocompatibility rationale.
- Accessories and components: List of all accessories with indication of which are covered by the submission.
- Previous generations: Information about predecessor devices if applicable.
Section 2: Information Supplied by the Manufacturer
Labeling Documentation Checklist
- Device labels: Artwork or specimens of all device labels.
- Packaging labels: All levels of packaging labeling.
- Instructions for Use (IFU): Complete IFU in all intended languages.
- Symbol glossary: Reference to standards for any symbols used (ISO 15223-1, etc.).
- UDI carrier: Evidence of compliant UDI placement and format.
- MDR Annex I compliance: Verification that labeling meets Chapter III requirements.
Section 3: Design and Manufacturing Information
Design Documentation Checklist
- Design and development plan: Evidence of planned approach to design activities.
- Design inputs: Documented requirements derived from intended purpose and user needs.
- Design outputs: Specifications, drawings, and manufacturing instructions.
- Design reviews: Evidence of formal reviews at appropriate stages.
- Design verification: Testing that outputs meet inputs, with protocols and reports.
- Design validation: Evidence that device meets user needs under actual or simulated conditions.
- Design transfer: Evidence that design was properly transferred to manufacturing.
- Design changes: Documentation of any changes made after initial design completion.
Manufacturing Documentation Checklist
- Manufacturing process description: Overview of all manufacturing steps.
- Manufacturing sites: List of all sites involved with addresses and activities performed.
- Critical suppliers: Identification of suppliers for critical components and materials.
- Process validations: Evidence that special processes are validated.
- In-process controls: Description of quality checks during manufacturing.
- Final product testing: Acceptance criteria and test methods.
Section 4: General Safety and Performance Requirements
GSPR Documentation Checklist
- GSPR checklist matrix: Tabular mapping of each Annex I requirement to evidence.
- Standards applied: List of harmonized standards with specific clauses applied.
- Common specifications: Application of any relevant common specifications.
- Non-applicable justifications: Rationale for any requirements deemed not applicable.
- Evidence references: Specific document numbers, sections, and page numbers for each requirement.
- Alternative solutions: For requirements not met via standards, description of alternative approaches.
Section 5: Benefit-Risk Analysis and Risk Management
Risk Management Checklist
- Risk management plan: Documented approach per ISO 14971.
- Hazard identification: Comprehensive list of potential hazards.
- Risk estimation: Probability and severity assessment for each hazard.
- Risk evaluation: Determination of acceptability against defined criteria.
- Risk control measures: Documented controls with verification of implementation.
- Residual risk evaluation: Assessment of remaining risks after controls.
- Overall residual risk: Judgment that total residual risk is acceptable.
- Benefit-risk determination: Explicit statement that benefits outweigh risks.
- Risk management report: Summary document with conclusions.
- Production and post-production information: Plan for ongoing risk monitoring.
Section 6: Product Verification and Validation
Testing Documentation Checklist
- Test protocols: Documented test methods with acceptance criteria.
- Test reports: Results with clear pass/fail determinations.
- Biocompatibility testing: ISO 10993 evaluation with biological testing as appropriate.
- Electrical safety: IEC 60601 series testing for electro-medical devices.
- EMC testing: Electromagnetic compatibility per IEC 60601-1-2.
- Sterilization validation: For sterile devices, complete validation per relevant standards.
- Packaging validation: Sterile barrier system testing, transport simulation.
- Shelf life/stability: Evidence supporting stated shelf life.
- Software verification and validation: For devices with software, per IEC 62304.
- Usability testing: Per IEC 62366-1, including formative and summative studies.
Section 7: Clinical Evaluation
Clinical Documentation Checklist
- Clinical evaluation plan: Documented approach to clinical evaluation.
- Clinical evaluation report: Complete CER per MEDDEV 2.7/1 Rev 4 structure.
- Literature search protocol: Documented search strategy with databases, terms, and criteria.
- Literature search results: Evidence of search execution with results.
- Literature appraisal: Critical analysis of relevant publications.
- Equivalence assessment: If claiming equivalence, demonstration of clinical, technical, and biological equivalence.
- Clinical data analysis: Synthesis of clinical evidence from all sources.
- Clinical investigation reports: If conducted, complete study documentation.
- PMCF plan: Post-Market Clinical Follow-up plan.
- Conclusions: Clear statement on conformity with GSPRs.
Section 8: Post-Market Surveillance
PMS Documentation Checklist
- PMS plan: Documented plan per Article 84.
- Data sources: Identification of all data sources to be monitored.
- Analysis methods: Description of how data will be analyzed.
- Reporting procedures: When and how PMS findings trigger updates or reports.
- PSUR/SSR procedures: Plans for Periodic Safety Update Reports or Summary Safety Reports.
- Trend reporting: Procedures for detecting and reporting trends per Article 88.
- Vigilance procedures: Serious incident and FSCA reporting processes.
Final Pre-Submission Checklist
- All documents have current version numbers and approval signatures.
- Cross-references between documents are verified and consistent.
- GSPR mapping is complete with specific evidence references.
- Clinical Evaluation Report meets MEDDEV 2.7/1 Rev 4 requirements.
- Risk management file is complete with benefit-risk determination.
- All labeling complies with MDR Annex I Chapter III requirements.
- PMS and PMCF plans are documented and resourced.
- Document index is complete and accurately reflects submission contents.
Common Notified Body Questions to Anticipate
Based on our experience, Notified Bodies frequently request clarification on:
- Equivalence claims: Be prepared to provide additional justification or clinical data if equivalence is challenged.
- Software classification: For devices with software, expect questions about SOUP (Software of Unknown Provenance) and cybersecurity.
- Biocompatibility rationale: Justify why certain tests were or were not performed.
- Clinical data gaps: Have a plan for generating additional clinical evidence if required.
- Supplier qualification: Be ready to provide evidence of critical supplier oversight.
- Design change history: Document why changes were made and their impact on safety and performance.
Working With Your Notified Body
A successful Notified Body relationship is built on thorough preparation and responsive communication. Consider these practices:
- Pre-submission meetings: Many Notified Bodies offer pre-submission consultations. Use these to clarify expectations and identify potential issues early.
- Response planning: Have resources allocated to respond to questions within the timeframes specified in your agreement.
- Question tracking: Maintain a log of all questions received and responses provided.
- Consistent contacts: Designate specific individuals as points of contact to ensure continuity.
Remember that Notified Bodies are assessing whether you have demonstrated conformity, not whether your device works. Your documentation must tell a complete, consistent, and compelling story of how each MDR requirement has been addressed.